Recessed elliptical-centric view-ordering for contrast-enhanced 3D MR angiography of the carotid arteries.

TitleRecessed elliptical-centric view-ordering for contrast-enhanced 3D MR angiography of the carotid arteries.
Publication TypeJournal Article
Year of Publication2002
AuthorsWatts R, Wang Y, Redd B, Winchester PA, Kent KC, Bush HL, Prince MR
JournalMagn Reson Med
Volume48
Issue3
Pagination419-24
Date Published2002 Sep
ISSN0740-3194
KeywordsCarotid Artery Diseases, Contrast Media, Fluoroscopy, Gadolinium DTPA, Humans, Imaging, Three-Dimensional, Magnetic Resonance Angiography
Abstract

Fast arterial-venous transit in the carotid arteries requires accurate, reliable timing of the acquisition to the bolus transit to maximize arterial signal and minimize venous artifacts. The rising edge of the bolus is not utilized in conventional elliptical-centric view-ordering because the critical k-space center must be acquired with full arterial enhancement. In this study, a recessed elliptical-centric view-ordering scheme is introduced in which the k-space center is acquired a few seconds following scan initiation. The recessed view-ordering is shown to be more robust to timing errors than the conventional scheme in a study of 37 patients.

DOI10.1002/mrm.10235
Alternate JournalMagn Reson Med
PubMed ID12210905
Grant ListR01HL60879 / HL / NHLBI NIH HHS / United States
R01HL62994 / HL / NHLBI NIH HHS / United States
Related Institute: 
MRI Research Institute (MRIRI)

Weill Cornell Medicine
Department of Radiology
525 East 68th Street New York, NY 10065