| Title | Optimal coil array design: the two-coil case. |
| Publication Type | Journal Article |
| Year of Publication | 2007 |
| Authors | Spincemaille P, Brown R, Qian Y, Wang Y |
| Journal | Magn Reson Imaging |
| Volume | 25 |
| Issue | 5 |
| Pagination | 671-7 |
| Date Published | 2007 Jun |
| ISSN | 0730-725X |
| Keywords | Computer Simulation, Contrast Media, Equipment Design, Gadolinium, Heterocyclic Compounds, Image Enhancement, Magnetic Resonance Imaging, Organometallic Compounds, Phantoms, Imaging |
| Abstract | The optimization problem for coil arrays is largely unsolved, even for the case of a two-coil system. This paper reports a systematic computer simulation to investigate the maximal achievable signal-to-noise ratio (SNR) with a two-coil receiver system where, using cancellation circuitry, mutual inductance is made zero. Both symmetrical and asymmetrical solutions with respect to two-coil geometry are considered. SNR is measured at a single point at a certain depth and also along a longitudinal or transverse line at the same depth. The conducting medium containing these regions of interest is assumed to be an infinite half space, an infinite cylinder or a finite sphere. The previous coil array design using a "magical" overlap only approximates the optimal solution for the infinite half space. For the infinite cylinder and the finite sphere, optimal solutions can be quite different from the "magical" overlap. |
| DOI | 10.1016/j.mri.2006.10.004 |
| Alternate Journal | Magn Reson Imaging |
| PubMed ID | 17540279 |
Related Institute: