Title | Clinical Quality Control of MRI Total Kidney Volume Measurements in Autosomal Dominant Polycystic Kidney Disease. |
Publication Type | Journal Article |
Year of Publication | 2023 |
Authors | Zhu C, Dev H, Sharbatdaran A, He X, Shimonov D, Chevalier JM, Blumenfeld JD, Wang Y, Teichman K, Shih G, Goel A, Prince MR |
Journal | Tomography |
Volume | 9 |
Issue | 4 |
Pagination | 1341-1355 |
Date Published | 2023 Jul 12 |
ISSN | 2379-139X |
Keywords | Humans, Kidney, Magnetic Resonance Imaging, Polycystic Kidney, Autosomal Dominant, Quality Control |
Abstract | Total kidney volume measured on MRI is an important biomarker for assessing the progression of autosomal dominant polycystic kidney disease and response to treatment. However, we have noticed that there can be substantial differences in the kidney volume measurements obtained from the various pulse sequences commonly included in an MRI exam. Here we examine kidney volume measurement variability among five commonly acquired MRI pulse sequences in abdominal MRI exams in 105 patients with ADPKD. Right and left kidney volumes were independently measured by three expert observers using model-assisted segmentation for axial T2, coronal T2, axial single-shot fast spin echo (SSFP), coronal SSFP, and axial 3D T1 images obtained on a single MRI from ADPKD patients. Outlier measurements were analyzed for data acquisition errors. Most of the outlier values (88%) were due to breathing during scanning causing slice misregistration with gaps or duplication of imaging slices (n = 35), slice misregistration from using multiple breath holds during acquisition (n = 25), composing of two overlapping acquisitions (n = 17), or kidneys not entirely within the field of view (n = 4). After excluding outlier measurements, the coefficient of variation among the five measurements decreased from 4.6% pre to 3.2%. Compared to the average of all sequences without errors, TKV measured on axial and coronal T2 weighted imaging were 1.2% and 1.8% greater, axial SSFP was 0.4% greater, coronal SSFP was 1.7% lower and axial T1 was 1.5% lower than the mean, indicating intrinsic measurement biases related to the different MRI contrast mechanisms. In conclusion, MRI data acquisition errors are common but can be identified using outlier analysis and excluded to improve organ volume measurement consistency. Bias toward larger volume measurements on T2 sequences and smaller volumes on axial T1 sequences can also be mitigated by averaging data from all error-free sequences acquired. |
DOI | 10.3390/tomography9040107 |
Alternate Journal | Tomography |
PubMed ID | 37489475 |
PubMed Central ID | PMC10366880 |
Related Institute:
MRI Research Institute (MRIRI)